"This study aimed to verify the efficacy of disinfection procedures to reduce Acinetobacter baumannii blaOXA-23 bacterial load in needleless connectors that had been experimentally contaminated" Biazus-Dalcin et al (2024).
Needleless connector disinfection procedure

Abstract:

Aim: This study aimed to verify the efficacy of disinfection procedures to reduce Acinetobacter baumannii blaOXA-23 bacterial load in needleless connectors that had been experimentally contaminated.

Methods: Two-way intermediate extender’s hub and needle-free valve were contaminated with Acinetobacter baumannii blaOXA-23. To disinfect them, the following procedures were carried out: sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% isopropyl alcohol single-use cap, with eight times friction for 10 s, followed by 5 s drying time. The statistical tests Kruskal-Wallis and post-hoc Conover were performed using MedCalc®.

Results: A total of 82 experiments were conducted. All tested disinfection procedures were efficacious in reducing the A. baumannii blaOXA-23 load. The 70% IPA single-use cap was found to be the best method for disinfecting the two-way intermediate extender’s hub (87.28%), while all the methods were efficacious for the disinfection of the needle-free valve (more than 90%). During the inoculation period, A. baumannii blaOXA-23 showed less adherence to the needle-free valve during the inoculation period, probably due to the device’s design.

Conclusion: The three tested disinfection procedures using sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% IPA single-use cap were found to be efficacious in reducing the bacterial load of A. baumanni blaOXA-23 in needleless connectors. Proper disinfection of needleless connectors is a crucial nursing practice to prevent bloodstream infections, as it significantly reduces the bacterial load present in the device.

Reference:

Biazus-Dalcin C, Sincero TCM, Zamparette CP, Tartari DC, de Souza S, Silva TL, Tomazoni A, Rocha PK. Efficacy of disinfection procedures to reduce Acinetobacter baumanii blaOXA-23 contamination rate of needleless connectors: an in-vitro study. Infect Prev Pract. 2023 Dec 1;6(1):100328. doi: 10.1016/j.infpip.2023.100328. PMID: 38226398; PMCID: PMC10788527.