Search
"Although features like DERS can help ensure safe medication administration, continuous improvement efforts to increase DERS compliance without increasing alert burden are needed to ensure that benefits of this technology are optimized" Hughes et al (2024).

Optimizing use of dose error reduction software on IV pumps

Abstract:

Background and objectives: Intravenous infusions have the potential to cause significant harm in patients and are associated with a high rate of adverse drug events and medication errors. Infusion pumps with dose error reduction software (DERS) can be used to reduce errors by establishing safe infusion parameters. In 2019, a quality improvement project was initiated with the aim to increase DERS compliance from 46% to 75% at our specialty institution by October 1, 2022.

Methods: An interdisciplinary group was tasked with improving compliance with DERS by identifying key drivers, including informed staff, engaged staff, and an accurate smart pump library. We used the Model for Improvement framework to guide this improvement project, and Plan-Do-Study-Act (PDSA) cycles were used to plan for interventions. PDSA cycles included drug library updates, education, and unit-level compliance reporting. Weekly average DERS compliance was monitored as the outcome measure, and weekly pump alerts per 100 infusions were monitored as a balancing measure; statistical process control charts were used to monitor measures from 2018 to 2022.

Results: Over the course of 25 months, 8 PDSA cycles resulted in 5 centerline improvements from a baseline mean of 46% to a final mean of 78%. Pump alerts per 100 infusions decreased from 15.9 to 6.4 with the first PDSA cycle and then continued to decrease to 3.9 with subsequent interventions.

Conclusions: Although features like DERS can help ensure safe medication administration, continuous improvement efforts to increase DERS compliance without increasing alert burden are needed to ensure that benefits of this technology are optimized.


Reference:

Hughes K, Cole M, Tims D, Wallach T, Spencer C, Page V, Robertson J, Hoffman JM. Optimizing the Use of Dose Error Reduction Software on Intravenous Infusion Pumps. Hosp Pediatr. 2024 May 8:e2023007385. doi: 10.1542/hpeds.2023-007385. Epub ahead of print. PMID: 38716570.