Search
"Tunneled catheter cuff extrusion is a frequent phenomenon. Catheter-related bloodstream infection, hypertension, and age of 61 to 95 years are high-risk factors for multiple episodes of cuff extrusion" Raza et al (2024).
Tunneled central venous catheter cuff extrusion

Abstract:

Objective: To identify underlying factors associated with single versus multiple expulsions of tunneled hemodialysis catheter cuffs and their associated complications.

Materials and methods: A multicenter, five-year retrospective data analysis was conducted on hemodialysis patients with tunneled catheters. Patient data included age, gender, BMI, and associated comorbidities. The type of procedure (new tunnel insertion vs. exchange), exit site infection, and catheter-related bloodstream infection (CRBSI) were also included.

Results: The data of 122 patients was analyzed. Seventy-eight patients had diabetes mellitus, 102 patients had hypertension, and 24 had ischemic heart disease. Fifty-one patients were on antiplatelet therapy, and nine patients were on oral steroids. The access site for 98 patients was the right internal jugular; for 19 patients, it was the left internal jugular; five patients had a femoral dialysis catheter. Patients were grouped into two categories: those with single cuff extrusion episodes and those with multiple cuff-out episodes. Sixty-three patients had one cuff extrusion, and 59 had multiple cuff-out episodes during the study period. Patients who had CRBSI and hypertension and were aged between 61 and 95 had multiple episodes of cuff extrusion and reached statistical significance. Exit-site infection, diabetes mellitus, antiplatelet therapy, gender, catheter site, and BMI had no statistical significance between the two groups.

Conclusion: Tunneled catheter cuff extrusion is a frequent phenomenon. Catheter-related bloodstream infection, hypertension, and age of 61 to 95 years are high-risk factors for multiple episodes of cuff extrusion.

Reference:

Raza H, Hashmi MN, Khan MA, Troncoso M, Christallone A, Alhammouri J, Hafez M, Khan MS, Hejaili F. Etiological Factors Affecting Single Versus Multiple Tunneled Hemodialysis Catheter Cuff Extrusions in the Saudi Population. Cureus. 2024 Jan 25;16(1):e52906. doi: 10.7759/cureus.52906. PMID: 38406109; PMCID: PMC10893556.