"The dual-lumen power injectable PICCs provided a reliable access for blood sample collection, decreasing the number of blind percutaneous venous punctures in allo-HSCT" Shih et al (2021).

Power injectable PICC for stem cell transplant

Abstract:

Aims and objectives: To explore whether dual-lumen power injectable peripherally inserted central catheters (PICCs) could be effectively and safely applied in allogeneic hematopoietic stem cell transplantation (allo-HSCT) and for serum cyclosporine level monitoring.

Background: Compared to conventional central venous access devices, PICC provides a feasible route not only for fluid infusion, but also for blood sample collection in patients undergoing oncological treatments.

Design: A prospective observational study was conducted according to the STROBE guidelines.

Methods: We prospectively evaluated the applications and complications of power injectable PICCs in 52 consecutive allo-HSCT recipients. We also compared the cyclosporine levels in 188 paired blood samples, simultaneously obtained via power injectable PICCs and percutaneous venous puncture, to investigate whether power injectable PICC is a feasible route for cyclosporine concentration monitoring in allo-HSCT.

Results: The median PICC placement duration was 29 days. The insertion-site blood oozing and central line-associated bloodstream infection rates were 36.5% (19/52) and 26.9% (14/52), respectively, indicating the feasibility of these PICCs for various applications in allo-HSCT. No power injectable PICC-related thrombotic adverse events were identified; 90.4% (47/52) of cases with power injectable PICC removal occurred because of lack of medical utility, suggesting that power injectable PICC-related complications were manageable. However, cyclosporine levels in samples obtained via these PICCs were significantly higher than those in samples obtained via percutaneous venous puncture (261.5 ± 139.2 vs. 232.4 ± 253.6 ng/ml; p = 0.019 [set 1]; 254.8 ± 89.3 vs. 225.1 ± 233.3 ng/ml; p<0.001 [set 2]; 283.6 ± 103.9 vs. 238.0 ± 254.7 ng/ml; p = 0.006 [set 3]; 291.0 ± 94.9 vs. 266.0 ± 274.7 ng/ml; p = 0.016 [set 4]).

Conclusion: The power injectable PICC is a feasible venous access device for allo-HSCT.

Relevance to clinical practice: The dual-lumen power injectable PICCs provided a reliable access for blood sample collection, decreasing the number of blind percutaneous venous punctures in allo-HSCT. However, its application in cyclosporine level monitoring needs further investigation.


Reference:

Shih YH, Teng CJ, Chen TC, Chang KH, Chen MH. Dual-lumen power injectable peripherally inserted central catheters in allogeneic hematopoietic stem cell transplantation: A prospective observational study. J Clin Nurs. 2021 Aug 29. doi: 10.1111/jocn.16020. Epub ahead of print. PMID: 34459038.