"To assess the safety and feasibility of intentionally positioning the catheter tip in the right atrium (RA) without an abutment during implantation of a totally implantable venous access port (TIVAP)" Park et al (2024).
Implantation of venous access port tip

Abstract:

Purpose: To assess the safety and feasibility of intentionally positioning the catheter tip in the right atrium (RA) without an abutment during implantation of a totally implantable venous access port (TIVAP).

Materials and methods: We enrolled 330 patients who had undergone TIVAP implantation between January and December 2016 and postoperative chest CT. The TIVAP was placed using the single-incision technique to access the axillary vein directly from the incision line. To position the catheter tip in the RA without abutment, blood return was checked before cutting. Catheter length and complications were evaluated by retrospectively reviewing medical images and records.

Results: All patients achieved successful catheter tip positioning without abutment or dysfunction. The median tip position was 15.3 mm distal to the cavoatrial junction (CAJ) on fluoroscopy and 6 mm distal to the CAJ on CT. Catheter tips migrated a median of 10.4 mm cephalically on CT compared to fluoroscopy. Thromboses were detected in the RA and superior vena cava in one patient each.

Conclusion: Intentional catheter tip positioning in the RA without abutment is a safe and feasible technique with a low incidence of thrombosis and no observed dysfunction.

Reference:

Park H, Seo TS, Song MG, Yang WJ. Feasibility and Safety of a Technique Intended to Place the Catheter Tip in the Right Atrium without Abutment Against the Cardiac Wall during Implantation of the Totally Implantable Venous Access Port. J Korean Soc Radiol. 2024 Jan;85(1):161-170. doi: 10.3348/jksr.2023.0061. Epub 2024 Jan 22. PMID: 38362390; PMCID: PMC10864156.