Search

This study gives an indication of what might be expected in critical care environments in terms of the volume of general infusion alarms and critical short half-life infusion alarms, as well as for clinician reaction times to critical short half-life infusion-interruption events” Waterson and Bedner (2019).

Abstract:

BACKGROUND: Alarm fatigue commonly leads to a reduced response to alarms. Appropriate and timely response to intravenous pump alarms is crucial to infusion continuity. The difficulty of filtering out critical short half-life infusion alarms from nonurgent alarms is a key challenge for risk management for clinicians. Critical care areas provide ample opportunities for intravenous medication error with the frequent administration of high-alert, critical short half-life infusions that require rigorous maintenance for continuity of delivery. Most serious medication errors in critical care occur during the execution of treatment, with performance-level failures outweighing rule-based or knowledge-based mistakes.

OBJECTIVE: One objective of this study was to establish baseline data for the types and frequency of alarms that critical care clinicians are exposed to from a variety of infusion devices, including both large volume pumps and syringe drivers. Another objective was to identify the volume of these alarms that specifically relate to critical short half-life infusions and to evaluate user response times to alarms from infusion devices delivering these particular infusions.

METHODS: The event logs of 1183 infusion pumps used in critical care environments and in general care areas within the European region were mined for a range of alarm states. The study then focused on a selection of infusion alarms from devices delivering critical short half-life infusions that would warrant rapid attention from clinicians in order to avoid potentially harmful prolonged infusion interruption. The reaction time of clinicians to infusion-interruption states and alarms for the selected critical short half-life infusions was then calculated.

RESULTS: Initial analysis showed a mean average of 4.50 alarms per infusion in the general critical care pump population as opposed to the whole hospital rate of 1.39. In the pediatric intensive care unit (PICU) group, the alarms per infusion value was significantly above the mean average for all critical care areas, with 8.61 alarms per infusion. Infusion-interruption of critical short half-life infusions was found to be a significant problem in all areas of the general critical care pump population, with a significant number of downstream (ie, vein and access) occlusion events noted. While the mean and median response times to critical short half-life infusion interruptions were generally within the half-lives of the selected medications, there was a high prevalence of outliers in terms of reaction times for all the critical short half-life infusions studied.

CONCLUSIONS: This study gives an indication of what might be expected in critical care environments in terms of the volume of general infusion alarms and critical short half-life infusion alarms, as well as for clinician reaction times to critical short half-life infusion-interruption events. This study also identifies potentially problematic areas of the hospital for alarm fatigue and for particular issues of infusion and infusion-line management. Application of the proposed protocols can help create benchmarks for pump alarm management and clinician reaction times. These protocols can be applied to studies on the impact of alarm fatigue and for the evaluation of protocols, infusion-monitoring strategies, and infusion pump-based medication safety software aimed at reducing alarm fatigue and ensuring the maintenance of critical short half-life infusions. Given the frequency of infusion alarms seen in this study, the risk of alarm fatigue due to the white noise of pump alarms present in critical care, to which clinicians are constantly exposed, is very high. Furthermore, the added difficulties of maintaining critical short half-life infusions, and other infusions in specialist areas, are made clear by the high ratio of downstream occlusion to infusion starts in the neonatal intensive care unit (NICU). The ability to quantitatively track the volume of alarms and clinician reaction times contributes to a greater understanding of the issues of alarm fatigue in intensive care units. This can be applied to clinical audit, can allow for targeted training to reduce nuisance alarms, and can aid in planning for improvement in the key area of maintenance of steady-state plasma levels of critical short half-life infusions. One clear conclusion is that the medication administration rights should be extended to include right maintenance and ensured delivery continuity of critical short half-life infusions.

You may also be interested in…

[rp4wp]
[button link=”https://humanfactors.jmir.org/2019/3/e14123/” color=”default”]Full Text[/button]
[spacer height=”20px”]



Reference:

Waterson, J. and Bedner, A. (2019) Types and Frequency of Infusion Pump Alarms and Infusion-Interruption to Infusion-Recovery Times for Critical Short Half-Life Infusions: Retrospective Data Analysis. JMIR Human Factors. 6(3), p.e14123. doi: 10.2196/14123.