Potential blood splatter from intravenous catheters with passive safety mechanism

0

“We report a comparative laboratory study between 2 peripheral intravenous catheters equipped with a passive fully automatic safety mechanism to assess generation of blood droplets during withdrawal” Rosenthal and Hughes (2015).

Reference:

Rosenthal, V.D. and Hughes, G. (2015) Fluid dispersal from safety cannulas: An in vitro comparative test. American Journal of Infection Control. 43(3), p.305-307.

Abstract:

We report a comparative laboratory study between 2 peripheral intravenous catheters equipped with a passive fully automatic safety mechanism to assess generation of blood droplets during withdrawal. One presented no fluid droplets, whereas the other presented droplets in 48% and 60% for the best and worst case, with analysis of variance showing positive effects on the number of droplets generated (P < .001). Safety devices can introduce hazards if health care workers are at risk from blood splatter.

Thank you to our partners for supporting IVTEAM

Share.

Comments are closed.

Free Email Updates
Join 5.5K IVTEAM members. Subscribe now and be the first to receive all the latest free updates from IVTEAM!
100% Privacy. We don't spam.