"All T34 and T34L (T60) ambulatory syringe pumps – check pumps before each use due to risk of under-infusion and no alarm (MDA/2020/007)" MHRA (2020).
Extract:

All T34 and T34L (T60) ambulatory syringe pumps – check pumps before each use due to risk of under-infusion and no alarm (MDA/2020/007). Manufactured by CME (a BD company) – updated advice to address ‘wear and tear’ of the syringe pump motor block which may lead to under-infusion.

Actions by user include:
  • Healthcare users should be aware of the updated Directions for Use (DFU) and actions in the manufacturer’s Field Safety Notice (MMS-19-1572) published in February 2020. This includes instructions to inspect the pump before each use

  • if you see white plastic debris on the lead screw, immediately stop using the pump and send it to BD or your chosen service provider for repair

  • a reminder that it’s vital to regularly check that an infusion is running as expected./li>

    Reference:

    MHRA (2020) All T34 and T34L (T60) ambulatory syringe pumps – check pumps before each use due to risk of under-infusion and no alarm (MDA/2020/007). February 25th.