Field safety notice Becton Dickinson recall extension set PCA-01 for PCS pump

0

The leakage may lead to an under infusion which would result in the patient not receiving the therapeutic level of medication to control their pain leading to a temporary harm.
NHS Supply Chain has supplied this product via our stock route as NPC FSB886” NHS Supply Chain (2017).

NHS Supply Chain report “Becton Dickinson is recalling specific lots of the Extension Set PCA-01 to be used with PCA Pump, as a result of customer complaints regarding leakages at the tubing to tubing connector both during set up and during clinical use, due to an inconsistent amount of solvent applied during the manual assembly process.
Impact:

The leakage may lead to an under infusion which would result in the patient not receiving the therapeutic level of medication to control their pain leading to a temporary harm. NHS Supply Chain has supplied this product via our stock route as NPC FSB886.

Full Alert

Thank you to our partners for supporting IVTEAM