Animas Corporation 2020 Insulin Infusion Pump: Class I Recall – False alarm or warning sound

0

Intravenous products: FDA report “Animas identified a component issue affecting Animas 2020 Infusion Insulin Pumps manufactured from March 1, 2012 to November 30, 2012. The component issue may trigger the pumps to sound a false alarm or warning related to one of the following:

  • “Loss of prime”
  • “Occlusion”
  • “No Cartridge detected”

If you receive any of these alarms, the pump may prompt you to complete the rewind, load and prime sequence to clear this alarm. Failure to follow the pump’s safety instructions and disconnect your infusion set from your body before the “rewind, load and prime” steps can lead to unintended delivery of insulin, placing you in danger of potential serious health risks, such as hypoglycemia.