To improve jugular central venous access device (CVAD) securement, prevent CVAD failure (composite: dislodgement, occlusion, breakage, local or bloodstream infection), and assess subsequent trial feasibility” Rickard et al (2016).

Abstract:

PURPOSE: To improve jugular central venous access device (CVAD) securement, prevent CVAD failure (composite: dislodgement, occlusion, breakage, local or bloodstream infection), and assess subsequent trial feasibility.

[ctt tweet=”ReTweet if useful… Improving jugular central vascular access device securement http://ctt.ec/CqT0Q+ @ivteam #ivteam” coverup=”CqT0Q”]

MATERIALS AND METHODS: Study design was a 4-arm, parallel, randomized, controlled, nonblinded, pilot trial. Patients received CVAD securement with (i) suture+bordered polyurethane (suture + BPU; control), (ii) suture+absorbent dressing (suture + AD), (iii) sutureless securement device+simple polyurethane (SSD+SPU), or (iv) tissue adhesive+simple polyurethane (TA+SPU). Midtrial, due to safety, the TA+SPU intervention was replaced with a suture + TA+SPU group.

RESULTS: A total of 221 patients were randomized with 2 postrandomization exclusions. Central venous access device failure was as follows: suture + BPU controls, 2 (4%) of 55 (0.52/1000 hours); suture + AD, 1 (2%) of 56 (0.26/1000 hours, P=.560); SSD+SPU, 4 (7%) of 55 (1.04/1000 hours, P=.417); TA+SPU, 4 (17%) of 23 (2.53/1000 hours, P=.049); and suture + TA+SPU, 0 (0%) of 30 (P=.263; intention-to-treat, log-rank tests). Central venous access device failure was predicted (P<.05) by baseline poor/fair skin integrity (hazard ratio, 9.8; 95% confidence interval, 1.2-79.9) or impaired mental state at CVAD removal (hazard ratio, 14.2; 95% confidence interval, 3.0-68.4).

CONCLUSIONS: Jugular CVAD securement is challenging in postcardiac surgical patients who are coagulopathic and mobilized early. TA+SPU was ineffective for CVAD securement and is not recommended. Suture + TA+SPU appeared promising, with zero CVAD failure observed. Future trials should resolve uncertainty about the comparative effect of suture + TA+SPU, suture + AD, and SSD+SPU vs suture + BPU.

Reference:

Rickard, C.M., Edwards, M., Spooner, A.J., Mihala, G., Marsh, N., Best, J., Wendt, T., Rapchuk, I., Gabriel, S., Thomson, B., Corley, A. and Fraser, J.F. (2016) A 4-arm randomized controlled pilot trial of innovative solutions for jugular central venous access device securement in 221 cardiac surgical patients. Journal of Critical Care. 36, p.35-42. [Epub ahead of print].

doi: 10.1016/j.jcrc.2016.06.006.