We report the first worldwide experience with continuous veno-venous hemodialysis (CVVHD) in children using the last generation Cardio-Renal Pediatric Dialysis Emergency Machine (CARPEDIEM)TM device” Vidal et al (2018).

Abstract:

We report the first worldwide experience with continuous veno-venous hemodialysis (CVVHD) in children using the last generation Cardio-Renal Pediatric Dialysis Emergency Machine (CARPEDIEM)TM device. Thirteen children received 1,008 h of CVVHD during 95 sessions, using a 0.15 (n = 7) or a 0.25 m2 (n = 6) hemofilter. The median weight was 3 kg (interquartile range [IQR] 2.5-6.2). In 10 patients, CVVHD was conducted using a 5 Fr double-lumen central vascular access, whereas in 3 children, bigger sizes were used (6.5 and 8 Ch). The median prescribed Qb was 17 mL/min (IQR 10-29.5), with a median Qd of 10 mL/min. Circuits were primed with 5% albumin in 12 out of 13 patients, using anticoagulation with heparin in all 13 cases. The median delivered/prescribed time ratio yielded a 100% result (95-100%). The most common cause for “downtime” was clotting that however occurred in only 3% of all treatments. Survivals at continuous renal replacement therapy discontinuation and ICU discharge were 100 and 69% respectively. The CARPEDIEMTM machine allowed successful delivery of diffusive blood purification modality to very small patients, using small catheters, no blood primes, and excellent concordance between delivered and prescribed treatment duration.

You may also be interested in…

[rp4wp]



Reference:

Vidal, E., Cocchi, E., Paglialonga, F., Ricci, Z., Garzotto, F., Peruzzi, L., Murer, L. and Ronco, C. (2018) Continuous Veno-Venous Hemodialysis Using the Cardio-Renal Pediatric Dialysis Emergency MachineTM: First Clinical Experiences. Blood Purification. October 31st. [epub ahead of print].

doi: 10.1159/000494437.